  1. 1. Turn on the autosampler. Screen will read "...STANDBY..." 2. Press "ENTER/PROGRAM" button. Use arrows to highlight "CONFIGURE". Press "ENTER/PROGRAM" button This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the "ENTER/PROGRAM" button.
    • 1. Turn on the autosampler. Screen will read "...STANDBY..."

    • 2. Press "ENTER/PROGRAM" button. Use arrows to highlight "CONFIGURE". Press "ENTER/PROGRAM" button

    • This will allow you to use the arrows to shuffle through the different configuration options. To select the one you want, press the "ENTER/PROGRAM" button.

    • 3. Use arrows to find "EVENT MARK". Press the "ENTER/PROGRAM" button.

  2. 4. Use arrow to select "CONTINUOUS SIGNAL". Press "ENTER/PROGRAM" button. 5. Use arrow to select "PUMP CYCLE". Press "ENTER/PROGRAM" button. It is critical this setting is configured in order for the sensor node to operate the autosampler!
    • 4. Use arrow to select "CONTINUOUS SIGNAL". Press "ENTER/PROGRAM" button.

    • 5. Use arrow to select "PUMP CYCLE". Press "ENTER/PROGRAM" button.

    • It is critical this setting is configured in order for the sensor node to operate the autosampler!

    • 6. Use arrows to find "EXIT CONFIGURATION." Press "ENTER/PROGRAM" button. This will bring up the screen "...STANDBY...

Ziellinie

Ein weiterer Nutzer hat diese Anleitung absolviert.

Brooke Mason

Mitglied seit: 29.05.2019

2.865 Reputation

17 Anleitungen geschrieben

Team

Open-Storm Mitglied von Open-Storm

Community

21 Mitglieder

27 Anleitungen geschrieben

0 Kommentare

Kommentar hinzufügen

Seitenaufrufe:

Letzten 24 Stunden: 0

Letzten 7 Tage: 0

Letzten 30 Tage: 0

Insgesamt: 71