Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement

Was du brauchst

  1. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement, Electrochemical Sensors: Schritt 1, Bild 1 von 1
    • Confirm which sensor you are intending to replace. Make note of it.

    • Nitrile gloves should be worn for the entire procedure.

  2. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 2, Bild 1 von 2 Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 2, Bild 2 von 2
    • Place the monitor face down on a clean surface.

    • Use the Phillips #0 screwdriver to remove the single 3 mm Phillips screw on the belt clip.

  3. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 3, Bild 1 von 1
    • Starting at the bottom of the case, pull the outer case away from the main body of the monitor.

    • Continue pulling out and away until the entire outer case has been removed.

  4. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 4, Bild 1 von 1
    • With the Phillips #0 screwdriver, remove all four 3 mm screws from the rear pump/sensor cover.

  5. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 5, Bild 1 von 1
    • Remove the outer pump/sensor housing cover and set it aside.

  6. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 6, Bild 1 von 1
    • Carefully grip and twist the outer edge of the sensor to remove the sensor from its insertion point.

    • Be careful not to twist too aggressively as this could damage the sensor attachment point in the monitor.

  7. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 7, Bild 1 von 3 Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 7, Bild 2 von 3 Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 7, Bild 3 von 3
    • Grip the replacement sensor on the outer edge.

    • Insert the new sensor in a straight downward orientation.

    • Be careful to line up the prongs with their insertion/attachment point in the monitor.

  8. Honeywell MultiRae Hydrogen Sulfide Electrochemical Sensors Replacement: Schritt 8, Bild 1 von 1
    • Reassemble the unit in reverse order.

    • After reassembly, power the unit on to ensure that the monitor recognizes the new sensor.

    • Calibrate the sensor per the manufacturer's instructions.

Abschluss

To reassemble your device, follow these instructions in reverse order.