Service Manual (New Generation) Starting from Serial number 3000
Table of contents
| 1. History | 2 |
|---|---|
| 2. Purpose | 2 |
| 3. General | 2 |
| 4. Packing and unpacking | 4 |
| 5. Block diagram of electronic parts | |
| 6. Bill of material for service parts | 5 |
| 7. How to perform the repair | 9 |
| 8. Explosion drawing part | 10 |
| 9. Explosion drawing part 2 | 11 |
| 10. Checklist for Function- and Safetytest | 12 |
Table Caption: This table outlines the sections of a service manual and their corresponding page numbers, covering topics from history and purpose to detailed repair instructions and testing procedures. It serves as a table of contents for a technical document likely related to electronic equipment maintenance.
1. Historie History
| Revision Date | Name | Change | |
|---|---|---|---|
| 17.09.2001 | M. Nunnenmacher | Initial Release | |
| 2 | 19.09.2001 | M. Nunnenmacher | Pos. 4.8 new partnumber 0.0013.104 |
| 3 | 15.10.2001 | M. Nunnenmacher | Pos. 1.9, 1.15, 1.16 new partnumber |
| 6.11.2001 | M. Nunnenmacher Update | reference run, new packaging | |
| 5 | 15.01.2002 | M Nunnenmacher | Update Pos 3.13,3.45,4.3, 4.5, Safetytest |
| 06.02.2002 | M. Nunnenmacher | Update Pos. 3.12 Labeled casting | |
| 27.03.2002 | M. Nunnenmacher | Update Pos. 1.1 part - number power cord | |
| 8 | 21.10.2002 | Update for all electronic parts; explosion drawings and reference run | |
| 02.06.2003 | S. Herr | Update Pos. 2.10 explosion drawing | |
| 10 | 05.04.2004 | S. Herr | Update Pos.1.6, function test |
Table Caption: This table tracks revisions to a document, noting the date, author, and the specific changes made, such as new part numbers, updates to existing parts, and modifications to drawings or tests.
2 Purpose
This service manual is to perform some repairs on ARTROMOT'-K3 products. Repair and maintenance work may only be carried out by authorised persons; as other wise all warranty services and liabilities shall be void. It is not allowed to utilize any other spare part which is not mentioned in chapter 6.
3. General
3.1 Electronic and cable
- Electronic devices as hand-held programming unit; motor electronics and spiral cable are not interchangeable with 3000.
- Make sure that the characteristic values of your power supply correspond to the voltage and frequency data indicated on the power adapter.
- Error possibilities; showed by the Hand-held programming unit:
DEVICE NEEDS REFERENCE RUN: The reference run was interrupted or the electronics were replaced.
- Perform a reference run. See chapter 7.4
POTENTIOMETER: Potentiometer shows incorrect value Check plug connection; replace potentiometer if persistent
POTENTIOM. CABLE: Spiral cable for knee case defective
- Replace spiral cable for knee case (Pos. 5.9), check plug connection
MOTOR DRIVER: Motor electronics defective Replace motor electronics (see chapter 7.3) if persistent
OVER CURRENT: Motor power to high Replace motor electronics (see chapter 7.3) if persistent
UNWANTED BUSY: Motor control unit is not answering Hand-held programming unit; motor electronics or spiral cable of the hand-held programming unit defective
CPM MEMORY RAM: Memory fault in the motor control unit
- Replace motor electronics (see chapter 7.3) if persistent
CPM MEMORY ROM: Memory fault in the motor control unit
- Replace motor electronics (see chapter 7.3) if persistent
COMMUNICATION: Communication problem between hand-held programming unit and the motor control unit Hand-held programming unit; motor electronics or spiral cable of the hand-held programming unit defective
CPM-DEVICE ERROR: General mistake in the motor control unit
Replace motor electronics (see chapter 7.3) if persistent
MOT. EN. TIMEOUT: Motor can not be activated Check motor electronics and plug connection for motor: Replace motor electronics (see chapter 7.3) and spiral cable of the hand-held programming unit if persistent
INVALID PARAMET: Motor control unit received a wrong parameter
- Hand-held programming unit; motor electronics or spiral cable of the hand-held programming unit defective
STOP RELEASE ERR: Stop line can not be released Remove hand-held programming unit or spiral cable (see chapter 7.2) if persistent
Motor electronics or connecting line to motor defective
MOTOR DISABLED: Motor is disconnected Motor electronics or connecting line to motor defective
MOTOR ERROR: Motor is not turning Check motor cable and motor. Replace motor electronics (see chapter 7.3) if persistent
RTC INIT ERROR: Hand-held programming unit defective Replace hand-held programming unit (see chapter 7.2) if persistent
RTC COM ERROR: Communication problem with handheld programming unit
Replace hand-held programming unit (see chapter 7.2) if persistent
RTC ERROR: Hand-held programming unit defective Replace hand-held programming unit (see chapter 7.2) if persistent
RANGE EXCEEDED: Potentiometer shows a value out of range
- Replace Potentiometer and/ or spiral cable
UNKNOWN CPM ERR : Unknown error in the motor control unit
- Replace motor electronics (see chapter 7.3) if persistent
UNDEFINED ERROR: Undefined error
- Hand-held programming unit, motor electronics or spiral cable of the hand-held programming unit defective
CPM 24V SUPPLY: 24V supply in motor control out of range
- Check power supply. Replace motor electronics (see chapter 7.3) if persistent
CPM 3.3V SUPPLY: 3.3V supply in motor control out of range
Replace motor electronics(see chapter 7.3) if persistent
HS 24V SUPPLY: 24V supply in hand-held programming out of range
Replace hand-held programming unit or spiral cable if persistent
HS 5V SUPPLY: 5V supply in hand-held programming unit out of range
Replace hand-held programming unit if persistent
HS 3.3V SUPPLY: 3.3V supply in hand-held programming unit out of range
Replace hand-held programming unit if persistent
3.2 Mechanics
- The Threaded spindle is not interchangeable with ARTROMOT'-K3 serial number 3000.
- Do not loosen the knurled handles completely for any adjustment. For operation or transport; make sure that the knurled handles are tight.
3.3 Others
- Do not clean the casing or the support with grease or oil.
- Do not utilize any solvent for cleaning the
4. Packing and unpacking
Preparations for transport the ARTROMOT"-K3 has to be done.