Table of contents
| 1. History | 2 |
|---|---|
| 2. Purpose | 2 |
| 3. General | 2 |
| 4. Packing and unpacking the ARTROMOT®-K1 | 5 |
| 5. Block diagram of the electronic parts ARTROMOT®-K1 | 6 |
| 6. Bill of material for service parts ARTROMOT®-K1 | 7 |
| 7. Figure for bill of material | 8 |
| 8. Special function Service menu for ARTROMOT®-K1Comfort and Standard version | 10 |
| 9. Special function Service menu for ARTROMOT®-K1Classic | 12 |
| 10. How to perform repairs | 14 |
| 11. Checklist of safety and function test ARTROMOT®-K1Comfort and Standard version | 15 |
| 12. Checklist of safety and function test ARTROMOT®-K1Classic | 16 |
Table Caption: This table outlines the structure and content of a technical manual for the ARTROMOT®-K1 device, detailing sections from its history and purpose to specific repair procedures, service menus, and safety checklists for different versions.
1. History
| Revision | Date | Name | Change |
|---|---|---|---|
| 1 | 10.02.2006 | S. Herr | Service Manual created |
| 2 | 14.07.2006 | S. Herr | Pos. 25, chapter 10 |
| 3 | 14.08.2006 | S.Herr | Pos. 20, Chapter 9, 12 |
Table Caption: This table details the revision history of a service manual, including the revision number, date, author, and specific changes made, often referencing positions and chapters within the document.
2. Purpose
The purpose of this Service Manual is to help you make simple repairs on the ARTROMOT®-K1. Only authorized staff may perform repairs and maintenance as the manufacturer's warranty and liability would otherwise be invalidated. Only original parts may be used for servicing in accordance with the attached spare parts list.
3. General
3.1 Electronics, connection cables
No plugs may be connected or disconnected while the unit is switched on. Always switch the ARTROMOT®-K1 off before connecting or disconnecting a plug.
The locks for spiral cable for the hand-held programming unit have to be closed at all times.
When you assembling with electronic parts make sure to use ESD (Electro Static Discharge) equipment.
Only original chip cards may be used. Insert the chip cards so that ARTROMOT® is visible.
If you have to exchange any of the printed circuit boards including the knee electronics, hand held programming unit, motor control or power supply electronics you have to perform a calibration.
Possible errors: Errors will be displayed on the handheld programming unit as follows shown:
ERROR XX
1 Potentiometer error:
Wrong angle information
- -> Check the femur settings
- -> Replace motor control (Pos. 25)
- -> Replace knee electronics (Pos. 5)
2 Failure at the potentiometer:
Connection to the potentiometer is interrupted
- -> Replace the spiral cable of the potentiometer (Pos. 24)
- -> Replace motor control (Pos. 25)
- -> Replace knee electronics (Pos. 5)
3 Motor driver error:
The motor driver IC reported an error
-> Replace motor control (Pos. 25)
4 Motor error:
The motor did not turn properly
- -> Replace motor control (Pos. 25)
- -> Replace the motor (Pos. 26)
(XX = Number of the error)
5 Motor over current:
The current for the motor exceeded the maximum limit
- -> Replace motor control (Pos. 25)
- -> Check the mechanics (Pos. 30)
- -> Replace the motor (Pos. 26)
6 Motor control error:
Internal error in the motor control.
- -> Replace motor control (Pos. 25)
7 Eprom access error:
Memory error in the access of the EPROM
- -> Replace the hand held programming unit (Pos. 20)
8 CPM ROM error:
Memory error in the motor control
- -> Replace motor control (Pos. 25)
9 Communication:
Communication to the motor control is not possible
- -> Check spiral cable and connector
- -> Replace motor control (Pos. 25)
- -> Replace the hand held programming unit (Pos. 20)
10 General error in the motor control:
Unknown error in the motor control
- -> Replace motor control (Pos. 25)
11 Motor enable timeout
Motor could not be enabled in time
- -> Replace motor control (Pos. 25)
12 Invalid parameter motor error:
Motor has received a invalid parameter
- -> Replace the hand held programming unit (Pos. 20)
- -> Replace motor control (Pos. 25)
13 Stop release error:
The motor could not be released
- -> Replace motor control (Pos. 25)
- -> Replace the hand held programming unit (Pos. 20)
14 Unexpected motor Stop:
- -> Check spiral cable and connector
- -> Replace motor control (Pos. 25)
15 Motor disabled:
Motor control disabled the motor.
- -> Replace motor control (Pos. 25)
16 Wrong command in the motor :
- -> Replace motor control (Pos. 25)
- -> Replace the hand held programming unit (Pos. 20)
17 5V supply error:
5V supply of motor control not sufficient
- -> Replace motor control (Pos. 25)
18 Initialise error real time clock:
- -> Replace the hand held programming unit (Pos. 20)
19 Communication error real time clock:
- -> Replace the hand held programming unit (Pos. 20)
20 Error real time clock :
- -> Replace the hand held programming unit (Pos. 20)
21 Range exceeded:
The measured angle is out of the range of motion
- -> Replace motor control (Pos. 25)
22 ROM error in the hand held programming unit:
Memory error in the hand held programming unit
- -> Replace the hand held programming unit (Pos. 20)
23 Invalid parameter:
Internal error in the hand held programming unit
- -> Replace the hand held programming unit (Pos. 20)
24 24V supply error motor control:
Error in the 24V supply in the motor control
- -> Replace the power supply electronics (Pos. 28)
- -> Replace motor control (Pos. 25)
25 Bus error:
Bus system error
- -> Replace the hand held programming unit (Pos. 20)
- -> Replace the spiral cable of the hand held programming unit
- -> Replace motor control (Pos. 25)
26 24V supply hand held programming unit:
24V supply of the hand held programming unit is
defective -> Replace the hand held programming unit (Pos. 20)
27 5V supply hand held programming unit:
5V supply of the hand held programming unit is defective
- -> Replace the hand held programming unit (Pos. 20)