GSI 38 Service Manual
Covers Welch Allyn TM262 Auto Tymp
Grason-Stadler, Inc:
GSI 38 Service Manual
Grason-Stadler, Inc. Westchester Drive Milford; NH 03055-3056 Telephone: 603-672-0470 Fax: 603-672-0487
1738-0110, Rev. 1 Printed 1993 Revised November 1993 Revised April 1995 May,
WARRANTY
We, Grason-Stadler, Inc. warrant that this product is free from defects in material a workmanship and, when properly installed and used, will perform in accordance with applicable specifications. If within one year after original shipment it is found not t meet this standard, it will be repaired, or at our option, replaced at no charge exc for transportation costs, when returned to an authorized Grason-Stadler service facilit If field service is requested, there will be no charge for labor or material; however, will be a charge for travel expense at the service center's current rate.
NOTE
Changes in the product not approved in writing by Grason-Stadler shall void this warranty. Grason-Stadler shall not be liable for any indirect, special or consequential damages, even if notice has been given in advance of the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO , AN Y IMPLIED WARRANTY OF MERCHANTABILITY OF FITNESS FOR A PARTICULAR PURPOSE. '
WARNING
The GSI 38 is designed to be used with a hospital grade outlet. Injury to personnel o damage to equipment can result when a three-prong to two-prong adapter is connected between the GSI 38 power plug and an AC outlet or extension cord. Additionally, those GSI 38 Auto Tymps that are equipped with power transformers, use a specific transformer (8000-0260, 800-0261, or 8000-0262) which should not be interchanged with any other transformer or supply.
The GSI 38 is a specifically calibrated device and the periodic service and adjustments which the instrument may require should be done only by an authorized Grason-Stadler service technician.
PRODUCT SPECIFICATION
GSI 38 PRODUCT SPECIFICATIONS (RI)
CATALOG LISTINGS
1738-9700
1738-9705
1738-9706
1738-9708
GSI 38 Auto Tymp, Vl, USA
Tymp + lpsi Reflex
GS I 38 Auto Tymp, Vl, Export
GSI 38 Auto Tymp, Vi, Export, 100V
GS I 38 Auto Tymp, V1, , Export, 240V
1738-9710
GSI 38 Auto Tymp, V2, USA
Tymp + lpsi/Contra Reflex
1738-9715
GS I 38 Auto Tymp , V2 , Export
1738-9716
GSI 38 Auto Tymp, V2, Export, 100V
1738-9716
GSI 38 Auto Tymp, V2, Export, 240V
1738-9720
GSI 38 Auto Tymp, V3, USA
Tymp +Ipsi/Contra + Aud.
1738-9725
GS I 38 Auto Tymp , V3 , Export
1738-9726
GSI 38 Auto Tymp, V3, Export, 100V
1738-9728
GSI 38 Auto Tymp, V3, Export, 240V
1738-9730
GSI 38 Auto Tymp, V4, USA
Tymp + lpsi + Aud.
1738-9735
GS I 38 Auto Tymp, V4 , Export
1738-9736
GSI 38 Auto Tymp, V4, Export, 100V
1738-9738
GSI 38 Auto Tymp, V4, Export, 240V
DESCRIPTION
The GSI 38 is an automatic acoustic admittance and reflex measurement system with a screening audiometer. The instrument consists of a probe and a desk-to-base interconnected via a cable. It is a combination of GSI 17 and GSI 37 plus acousti reflex testing capabilities, both ipsilateral and contralateral. The audiometric port a single channel pure tone audiometer, with a pair of air-conduction earphones. Int h tympanometric mode, the measurement of acoustic admittance is corrected for the acoustic admittance of the external auditory meatius, and the ear canal volume is displayed. Also a test cavity assembly is provided.
INSTRUCTION MANUAL
The GSI 38 is supplied with an instruction manual which meets the requirements of the applicable standards.
STANDARDS
ANSI S3.39 - 1987
ANSI S3.6 - 1989
IEC 1027-l 991
IS0 645
IEC 601-l
Aural Acoustic Impedance and Admittance Standard Type 3 (Middle Ear)
Audiometric Standard Type 4 (Audiometer)
Aural Acoustic Impedance/Admittance Type 3
Type 4 (Audiometer)
Medical Electrical Equipment General Requirements For Safety
Designed to Meet International Standards Organization requirements fo r ElectroMagnetic Compatibility and Safety
ACOUSTIC ADMITTANCE MEASUREMENT SYSTEM - AUTO TYMP
Probe Signals
Fre q uenc y :
226 Hz, +/- 3%
Sound Pressure Level :
85.5 dB SPL , +/- 2.0 dB , Measured in a 2. 0 cm 3 coupler (ANS I Spe c 9 0 dB SPL Max)
Harmonic Distortion:
< 5%
Acoustic Admittanc e Range
0.0 to 1. 5 cm 3 or 0.0 to 3.0 cm 3 - selected automatically for compensated tympanogram.
0.2 to 6. 0 cm 3 - ECV/cavity limits for initiating pressurization (starting test)
The 0.2 to 6. 0 cm 3 is the sum of the ear canal volume and the middle ear admittance.
(Full pressure sweep for 6 cc up to 7000 ft. altitude with no leak)
+/- 5% or +/- 0.1 cm 3 whichever is greater
Tvmpano g ram Gradient:
Pressure interval at 50% of compensated admittance peak.
Acoustic Admittance Accuracy
Pneumatic Svstem
Pressure Range (corrected for altitude):
- 200 to - 400 dapa (1 dapa = 1.02 mm water)
Pressure sweeps to at least -100 dapa if tymps complete otherwise sweeps to -400 daPa
Pressure Accuracv:
+/- 15% or +/- 10 daPa whichever is greater.
Rate of Sweep:
600 daPa/sec - for tymp slopes of <0.2 ml per 2 4 daPa
200 daPa/sec - for tymp slopes =,> 0.2 ml per 2 4 daPa
Test Time
Approximately 1 sec (high relative tymp will take longer).
Direction o f Sweep :
Positive to Negative
ACOUSTIC - REFLEX ACTIVATING STIMULUS SYSTEM
Stimulus Frequencies:
500, 1000, 2000, and 4000 Hz
Stimulus Freauencv Accuracv: +/- 3%
Stimulus Total Harmonic Distortion:
Stimulus Output Levels:
<5% measured acousticall y except at 4 khz where speaker driving signa l is measured
IPSI:
500, 4000 Hz
80, 90, 10 0 dB HL
1000, 2000 Hz
85, 95, 10
5 dB HL
(Time multiplexed with probe tone, 106 ms On/53 ms Off Cycle)
CONTRA: 500, 1000, 2000 and 4000 HZ 90, 100, 110 db HL
(Steady tone, non multiplexed)
Equivalent Hearing Threshold Levels, in DB SPL
**Table Caption: This table presents auditory evoked potential (AEP) thresholds for a transducer, TvPe, across