GSI 38 Service Manual

96 Seiten · 2.26 MB · PDF

Covers Welch Allyn TM262 Auto Tymp

Grason-Stadler, Inc:

GSI 38 Service Manual

Grason-Stadler, Inc. Westchester Drive Milford; NH 03055-3056 Telephone: 603-672-0470 Fax: 603-672-0487

1738-0110, Rev. 1 Printed 1993 Revised November 1993 Revised April 1995 May,

WARRANTY

We, Grason-Stadler, Inc. warrant that this product is free from defects in material a workmanship and, when properly installed and used, will perform in accordance with applicable specifications. If within one year after original shipment it is found not t meet this standard, it will be repaired, or at our option, replaced at no charge exc for transportation costs, when returned to an authorized Grason-Stadler service facilit If field service is requested, there will be no charge for labor or material; however, will be a charge for travel expense at the service center's current rate.

NOTE

Changes in the product not approved in writing by Grason-Stadler shall void this warranty. Grason-Stadler shall not be liable for any indirect, special or consequential damages, even if notice has been given in advance of the possibility of such damages.

THIS WARRANTY IS IN LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO , AN Y IMPLIED WARRANTY OF MERCHANTABILITY OF FITNESS FOR A PARTICULAR PURPOSE. '

WARNING

The GSI 38 is designed to be used with a hospital grade outlet. Injury to personnel o damage to equipment can result when a three-prong to two-prong adapter is connected between the GSI 38 power plug and an AC outlet or extension cord. Additionally, those GSI 38 Auto Tymps that are equipped with power transformers, use a specific transformer (8000-0260, 800-0261, or 8000-0262) which should not be interchanged with any other transformer or supply.

The GSI 38 is a specifically calibrated device and the periodic service and adjustments which the instrument may require should be done only by an authorized Grason-Stadler service technician.

PRODUCT SPECIFICATION

GSI 38 PRODUCT SPECIFICATIONS (RI)

CATALOG LISTINGS

1738-9700

1738-9705

1738-9706

1738-9708

GSI 38 Auto Tymp, Vl, USA

Tymp + lpsi Reflex

GS I 38 Auto Tymp, Vl, Export

GSI 38 Auto Tymp, Vi, Export, 100V

GS I 38 Auto Tymp, V1, , Export, 240V

1738-9710

GSI 38 Auto Tymp, V2, USA

Tymp + lpsi/Contra Reflex

1738-9715

GS I 38 Auto Tymp , V2 , Export

1738-9716

GSI 38 Auto Tymp, V2, Export, 100V

1738-9716

GSI 38 Auto Tymp, V2, Export, 240V

1738-9720

GSI 38 Auto Tymp, V3, USA

Tymp +Ipsi/Contra + Aud.

1738-9725

GS I 38 Auto Tymp , V3 , Export

1738-9726

GSI 38 Auto Tymp, V3, Export, 100V

1738-9728

GSI 38 Auto Tymp, V3, Export, 240V

1738-9730

GSI 38 Auto Tymp, V4, USA

Tymp + lpsi + Aud.

1738-9735

GS I 38 Auto Tymp, V4 , Export

1738-9736

GSI 38 Auto Tymp, V4, Export, 100V

1738-9738

GSI 38 Auto Tymp, V4, Export, 240V

DESCRIPTION

The GSI 38 is an automatic acoustic admittance and reflex measurement system with a screening audiometer. The instrument consists of a probe and a desk-to-base interconnected via a cable. It is a combination of GSI 17 and GSI 37 plus acousti reflex testing capabilities, both ipsilateral and contralateral. The audiometric port a single channel pure tone audiometer, with a pair of air-conduction earphones. Int h tympanometric mode, the measurement of acoustic admittance is corrected for the acoustic admittance of the external auditory meatius, and the ear canal volume is displayed. Also a test cavity assembly is provided.

INSTRUCTION MANUAL

The GSI 38 is supplied with an instruction manual which meets the requirements of the applicable standards.

STANDARDS

ANSI S3.39 - 1987

ANSI S3.6 - 1989

IEC 1027-l 991

IS0 645

IEC 601-l

Aural Acoustic Impedance and Admittance Standard Type 3 (Middle Ear)

Audiometric Standard Type 4 (Audiometer)

Aural Acoustic Impedance/Admittance Type 3

Type 4 (Audiometer)

Medical Electrical Equipment General Requirements For Safety

Designed to Meet International Standards Organization requirements fo r ElectroMagnetic Compatibility and Safety

ACOUSTIC ADMITTANCE MEASUREMENT SYSTEM - AUTO TYMP

Probe Signals

Fre q uenc y :

226 Hz, +/- 3%

Sound Pressure Level :

85.5 dB SPL , +/- 2.0 dB , Measured in a 2. 0 cm 3 coupler (ANS I Spe c 9 0 dB SPL Max)

Harmonic Distortion:

< 5%

Acoustic Admittanc e Range

0.0 to 1. 5 cm 3 or 0.0 to 3.0 cm 3 - selected automatically for compensated tympanogram.

0.2 to 6. 0 cm 3 - ECV/cavity limits for initiating pressurization (starting test)

The 0.2 to 6. 0 cm 3 is the sum of the ear canal volume and the middle ear admittance.

(Full pressure sweep for 6 cc up to 7000 ft. altitude with no leak)

+/- 5% or +/- 0.1 cm 3 whichever is greater

Tvmpano g ram Gradient:

Pressure interval at 50% of compensated admittance peak.

Acoustic Admittance Accuracy

Pneumatic Svstem

Pressure Range (corrected for altitude):

  • 200 to - 400 dapa (1 dapa = 1.02 mm water)

Pressure sweeps to at least -100 dapa if tymps complete otherwise sweeps to -400 daPa

Pressure Accuracv:

+/- 15% or +/- 10 daPa whichever is greater.

Rate of Sweep:

600 daPa/sec - for tymp slopes of <0.2 ml per 2 4 daPa

200 daPa/sec - for tymp slopes =,> 0.2 ml per 2 4 daPa

Test Time

Approximately 1 sec (high relative tymp will take longer).

Direction o f Sweep :

Positive to Negative

ACOUSTIC - REFLEX ACTIVATING STIMULUS SYSTEM

Stimulus Frequencies:

500, 1000, 2000, and 4000 Hz

Stimulus Freauencv Accuracv: +/- 3%

Stimulus Total Harmonic Distortion:

Stimulus Output Levels:

<5% measured acousticall y except at 4 khz where speaker driving signa l is measured

IPSI:

500, 4000 Hz

80, 90, 10 0 dB HL

1000, 2000 Hz

85, 95, 10

5 dB HL

(Time multiplexed with probe tone, 106 ms On/53 ms Off Cycle)

CONTRA: 500, 1000, 2000 and 4000 HZ 90, 100, 110 db HL

(Steady tone, non multiplexed)

Equivalent Hearing Threshold Levels, in DB SPL

**Table Caption: This table presents auditory evoked potential (AEP) thresholds for a transducer, TvPe, across