The Midmark M11 Steam Sterilizer is designed to be safe, reliable and easy to use. With one of the largest chambers of any tabletop autoclave on the market, the M11 can optimize your supply of available sterile instruments by reducing the time and effort required for sterilization.

1 Fragen Alle anzeigen

water leaks after cycle

water leaks after cycle completes

Diese Frage beantworten Ich habe das gleiche Problem

Ist dies eine gute Frage?

Bewertung 0
Einen Kommentar hinzufügen

2 Antworten

A small amount of water is normal after a cycle. The door will open to dry, and some steam will condense and drip on the counter. A large amount of water usually indicates a bad door gasket or door dam.

Paul Olson

Rescue Biomedical Services

https://rescuebiomed.com/autoclaves-and-...

War diese Antwort hilfreich?

Bewertung 0
Einen Kommentar hinzufügen

Is the half moon dam gasket in place on the bottom of the door?