Zum Hauptinhalt wechseln

ventilator ltv 950 not working. power is on completed post test.

ventilator ltv 950 not working. power is on completed post test. but on vent check mode always. its not working normally. ventilator enters into vent check mode automatically on switching on.

Diese Frage beantworten Ich habe das gleiche Problem

Ist dies eine gute Frage?

Bewertung 0
Einen Kommentar hinzufügen

2 Antworten

Since this is a medical device, I think it will be hard to find information.

Based on your question, it is not getting feedback from the device that checks the vent and just hangs there. Similar to what Windows OS will do if it doesn't get feedback from a USB port.

Could be caused by a blockage. Any other cause is probably not user repairable.

War diese Antwort hilfreich?

Bewertung 0
Einen Kommentar hinzufügen

you may have 2 situations, one can be select button is stuck and need to replace front panel, secondly check dip switch #5 to make sure it is off position (Switch toward outside ventilator). also #6&8 dip need to be off

War diese Antwort hilfreich?

Bewertung 0
Einen Kommentar hinzufügen

Antwort hinzufügen

RAJ KOTTAK wird auf ewig dankbar sein.
Seitenaufrufe:

Letzte 24 Stunden: 0

Letzte 7 Tage: 1

Letzte 30 Tage: 1

Insgesamt: 27