Hier findest du Antworten in diesem Dokument
Wir haben alle Seiten unter 174 gelesen, damit du es nicht tun musst. Frag einfach und erhalte jetzt deine Antwort.
AV-S Ventilator
AV-S Ventilator
Introduction
Partnership for Life
AV-S Ventilator
Product Overview
Introduction
Characteristics
Configuration
Sub-assemblies
AV-S Ventilator
Operation
Ventilator control settings
Operation workshop
Pre-operative checkout procedure
AV-S Ventilator
TECHNICAL DESCRIPTION
AV-S Pneumatic description.
AV-S Electrical description.
AV-S Ventilator
Practical:
Disassembly.
Planned maintenance.
Test and calibration procedures.
Fault finding.
AV S ventilator
Product Overview
Partnership for Life
Aims and Objectives
Aim:
The aim of this module is to give the student a full understanding of the features and characteristics of the AV-S ventilator.
Objective:
♦ By the end of this module the student will be able to:
Describe the features of the AV-S.
Recognise user controls.
Recognise AV-S configuration.
AV-S Ventilator
Pneumatically driven.
Software controlled.
Multi-mode ventilator.
Time-cycled, volume/pressure controlled and pressure limited.
Compliance & fresh gas compensated.
User adjustable sigh & inspiratory pause.
AV-S monitoring
Integrated monitoring:
Inspired oxygen measurement.
Measured from inspiratory limb.
Airway pressure measurement.
Measured in the expiratory limb.
Tidal/Minute volume measurement.
Dual spirometry system.
AV-S features
Large 8.4" high definition colour screen.
Touch screen.
Rotary control dial.
Single/ Dual waveform display.
Designed to integrate with:
PrimaSP; SP2 and SP3.
A100SP; A200SP.
AV-S features
Print outs & interfacing:
USB, Serial Port, Analogue outputs, HPPCL4 printer output.
30 minutes battery back up as standard:
Optional 60 minute battery available.
AV-S features
Membrane keys
LED power indicators:
Yellow when connected / battery
charging
LED on/off indicator.
Green when switched on
Menu selection key

AV-S
AV-S
On/Off Switch
Alarm silence/mute key.
Rotary control dial.
Partnership for Life
Modes of ventilation
Mandatory modes:
Volume Mode.
Pressure Mode.
Spontaneous breathing support modes:
SIMV.
SMMV.
Pressure Support.
Partnership for Life
Modes of ventilation
Volume Controlled Ventilation
◆ AV-S delivers a mandatory set volume of gas at fixed breath intervals.
The Patient is making no respiratory effort.
♦ AV-S is volume limited and time cycled.
♦ Tidal volume setting range, 20 to 1600ml.
Inspiration terminated after a preset time.
Modes of ventilation
Airway Flow and Pressure Patterns

Modes of ventilation
Volume ventilation parameter limits:
Tidal volume adjustable from 20 – 1600 mL
Rate adjustable from 4 – 100 bpm
I:E ratio adjustable from 1:0.3 – 1:8
♦ Peep 'Off or adjustable from 4 – 20 cmH $ _{2} $O
Inspiratory pressure limit from 10 to 80 cmH $ _{2} $O
Inspiratory pause 0-60%
Does not affect I:E ratio
Sigh 1.5 x Set Vt
Adjustable 1:10 to 1:100 breaths
Partnership for Life
1.87.xx display

Litres 10 0.8 0.6 0.4 0.2 0.0 -10 cmH₂O 60 50 40 30 20 10 0 0 0 0 0 0 0 0 0 0 V_T SET mL 600 mL SET bpm 10 SET: E