daeg SpO2 Analyzer User and Service Manual

Technical reference 42 Seiten · 1.09 MB · PDF

daeg $ \times $ SPO2 Analyzer User and Service Manual

METRON

Metron Offices

USA

1345 Monroe NW, Suite 255A

Grand Rapids, MI 49505

Phone: (+1) 888 863-8766

Fax: (+1) 616 454-3350

E-mail: support@metron.us

FRANCE

30, rue Paul Claude

F-91000 Evry

Phone: (+33) 1 6078 8899

Fax: (+33) 1 6078 6839

E-mail: info@metron.fr

GERMANY

Römerstrasse 22

D-63741 Aschaffenburg

Phone: (+49) 6021 447 9807

Fax: (+49) 6021 447 9808

E-mail: proteges@t-online.de

NORWAY

Vegamot 8

N-7048 Trondheim

Phone: (+47) 7395 4700

Fax: (+47) 7395 4701

E-mail: support@metron.no

Disclaimer

METRON provides this publication as is without warranty of any kind, either express or implied, including but not limited to the implied warranties of merchantability or fitness for any particular purpose. Further, METRON reserves the right to revise this publication and to make changes from time to time to the content hereof, without obligation to METRON or its local representatives to notify any person of such revision or changes. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.

Limited Warranty

METRON warrants that the daeg SPO2 Analyzer will substantially conform to published specifications and to the documentation, provided that it is used for the purpose for which it was designed. METRON will, for a period of twelve (12) months from date of purchase, replace or repair any defective analyzer, if the fault is due to a manufacturing defect. In no event will METRON or its local representatives be liable for direct, indirect, special, incidental, or consequential damages arising out of the use of or inability to use the daeg SPO2 Analyzer, even if advised of the possibility of such damages. METRON or its local representatives are not responsible for any costs, loss of profits, loss of data, or claims by third parties due to use of, or inability to use the daeg SPO2 Analyzer. Neither METRON nor its local representatives will accept, nor be bound by any other form of guarantee concerning the daeg SPO2 Analyzer other than this guarantee. Some jurisdictions do not allow disclaimers of expressed or implied warranties in certain transactions; therefore, this statement may not apply to you.

Table of Contents

Metron Offices.....2
Table of Contents.....3
daeg SpO2 Analyzer Specifications.....5
daeg Quick Start.....7
Unit Setup.....9
Unpacking.....9
Battery Installation.....9
Operation.....11
Power up Screen.....11
Main Simulation Screens.....11
Function Selection Menu.....12
Alarm Limits.....12
Remote Control.....13
Probe Analyzer.....13
Simulation Setup.....14
Print Options.....15
Troubleshooting.....17
Calibration.....19
Schematics.....21

This page was intentionally left blank.

daeg ✗ SpO2 Analyzer Specifications

SPO2 SIMULATION

GENERAL

Factory installed, downloadable and user programmable R-curves, patient conditions and templates.

  • Probe signal and LED test x

ACTIVE PROBE SPO2 SIMULATION

O2 saturation range: 30 - 100 %

Simulation resolution: 1% SpO2%

Simulation accuracy:

± 0.5 SpO2% 65-100% SpO2%

$ \pm 1 $ SpO2% 30-64% SpO2%

  • Simulation repeatability:

±0.2 SpO2%

Simulation rate: 30 - 300 bpm, res. 5 bpm (sync. to ECG for "C-lock" testing)

Rate accuracy: $ \pm 0.1\% $

Pulse amplitude: 0 - 100% of nominal pleth amplitude.

  • Amplitude accuracy: ≥30% amplitude ±1%. 1-29% amplitude ±5%.

Low, medium and high

  • Light repeater bandwidth: DC - 1MHz

  • Optical slew rate: 50 mcd/ $ \mu $s

  • Light repeater dynamic range: 1mcd + 13 dB - 60 dB

  • Receiver optical 3 dB bandwidth: 600 - 1050 nm.

INSTALLED R-CURVES

BCI International

  • Criticare

  • Masimo

  • Nellcor

Novametrix

  • Philips (Agilent)

  • Ohmeda

INSTALLED PATIENT CONDITIONS

  • Normal patient

  • Geriatric patient

Obese patient

Patient with weak pulse

Bradycardia

Tachycardia

ARTIFACTS (ACTIVE PROBE ONLY)

  • Motion: 0 - 4Hz

Ambient sunlight: Ambient, Sun

  • Interfering light: 50Hz, 60Hz

PROBE LED OPTICAL FREQUENCY

  • Frequency range: 550 - 1050 nm

  • Resolution: 1nm

  • Accuracy: $ \pm2\% $ of range

PROBE TEST

Available with release of Probe Test Adaptor Box and firmware update.

ECG SIMULATION - GENERAL

The data below is related both to factory installed and downloadable waveforms.

Lead configuration:

5 lead RL, RA, LA, LL, V1-6

Low level output amplitude: 0.2-2mV, resolution 100 mV

Low level output impedance: 1000 ohm to RL

High level output amplitude: 0.5V/mV of low level.

High level output impedance: 50 ohm

System amplitude accuracy: $ \pm2\% $ @1mV

System rate accuracy: $ \pm 0.1\% $

System time resolution: 10 µs (100 kHz maximum update rate)

ECG OUTPUT

The data below is related to factory installed waveform types.

NORMAL ECG SINUS

  • 30 - 300 BPM, resolution 1 BPM

GENERAL INFORMATION

POWER SUPPLY

  • Power sources: 4 standard AA/LR6/MN-1500 cells. 230VAC/115VAC to 9VDC battery eliminator.

Power consumption:

200 mA w/o display backlight.

400 mA w/full backlight

400 mA w/full backlight.

Battery lifetime: 12 h for alkaline cells without backlight.

HOUSING

High impact plastic case.

DIMENSIONS

  • Height: 237 mm/9.2"

  • Width: 122 mm/4.8"

42 mm/1.6"

WEIGHT

With battery: 0.6 kg/1.3 lb.

TEMPERATURE

  • Storage temperature: 0/32 to +50/122 °C/F

  • Operating temperature: +15/59 to +35/95°C/F

HUMIDITY

Operating humidity: 10% - 80% (non-condensing)

CONNECTORS

Graphical LCD, with backlighting.

  • 9V DC power inlet

DISPLAY

RS-232/C for PC or printer communication (9-pin D-sub male)

  • 1 probe connector (Redel 8-pin).

  • 5 low level ECG outputs RA, RL, LA, LL and V1 (AHA color coded safety connector)

1 high level ECG output (standard phono jack).

daeg SpO2 Analyzer

Ordering Information

Order no:

12700: daeg SpO2 Analyzer

Accessories:

17021: Battery Eliminator 240V or

17027: Battery Eliminator 115V

17030: Battery Package. 4-cells

12710: User and Service Manual

Options:

11150: Carrying case

12711: Cable Test Adapter Box

12712: PRO-Soft daeg

12713: PRO-Soft daeg, demo

12714: User Manual PRO-Soft daeg

17024: Universal Banana Adapter