JULIAN TECHNICAL SERVICE BULLETIN #27
Product:
Julian
Date:
March 23, 2001
Re:
Corrective action - replacement of Man/Spont-Auto switchover diaphragm located in valve plate of the breathing system.
Reference Doc:
03/01 Maxeiner No.: 27 Released 3/9/2001
Reason/Rational:
To enhance the reliability of the Man/Spont switchover valve.
Solution/Procedure:
Replacement of individual diaphragms of diaphragm assembly for Man/Spont-Auto switchover.
Notice: When fitting the diaphragm(s), pay special attention to the correct fitting position (see drawing). Secure the screw with Loctite 221.
Devices affected:
All units manufactured prior to December 2000. (serial numbers prior to ARPN-xxxx)
Timing/When:
During next repair or Preventative Maintenance Certification visit, but action must be completed by March 2003 .
Cost:
Parts will be supplied free of charge up to the completion date of March 2003.
Time Required:
Without wait or travel times:
Fax: (215) 723-5935
DrägerService is a division of
Draeger Medical, Inc.
RETURN TO CD-ROM TABLE OF CONTENTS RETURN TO TECHNICAL SERVICE BULLETIN TABLE OF CONTENTS
Disassembly and assembly of the diaphragm assembly: 10 min Function test: 15 minutes
Total: 25 minutes
Ordering Info/Parts: M32514
- (2x for each breathing system) Man/Spont diaphragm. Diaphragms can be ordered free-of-charge through Customer Service.
Checking/Testing:
Functional test: Carry out power on test, check ventilation parameters according to section 4.18- Ventilation Modes, of the Safety and Function test.
Distribution:
All Authorized Service Organizations, DragerService.
If you have any questions, please contact Technical Support by phone at 1-800-543-5047 or by fax at 1-215-721-5789.
DrägerService 3122 Commerce Drive Telford, PA 18969 Tel: (215) 721-5402 (800) 543-5047
Fax: (215) 723-5935
Fax: (215) 723-5935
Web: www.draegermedical.com